Phase encoding scheme handling
==============================
From version 3.0RC1 onwards, *MRtrix3* is capable of importing information from
DICOM relating to the phase encoding of the acquired images, and encoding this
information within key-value fields inside an image header. This information can
then later be used by the ``dwifslpreproc`` script, specifically using its
``-rpe_header`` command-line option, to automatically set up and execute FSL's
``topup`` and ``eddy`` commands without requiring explicit input from the user
regarding the phase encoding design of the imaging experiment. This page
explains how this information is encoded and manipulated by *MRtrix3*.
.. NOTE::
Due to variations in sequences, acquisition protocols, DICOM encoding, and
image manipulation operations, *MRtrix3* cannot be guaranteed to obtain or
retain the phase encoding information correctly in all circumstances. Manual
inspection of these data and/or the outcomes of image processing when making
use of these data is advocated. If circumstances leading to incorrect import
or encoding of this information in *MRtrix3* is reproducible, please report
this to the *MRtrix3* developers so that the software can be updated to
support such data.
Phase encoding information storage
----------------------------------
The phase encoding information for a particular image file can be stored in one
of two ways.
- The most convenient of these is storage of (one or more) key-value field(s)
encapsulated within the image header, just as can be used for
the diffusion gradient scheme. This ensures that the information
is retained through image processing, as each *MRtrix3* command passes the
header entries of the input image through to the output image.
- Alternatively, this information can be stored within a JSON file that
accompanies the relevant image file(s). This information would then typically
be imported / exported using the ``-json_import`` and ``-json_export`` options
in ``mrconvert``.
Precisely *how* this phase encoding information is *encoded* however depends on
the nature of the phase encoding information for that image; specifically,
whether the phase encoding information is *identical* for all volumes within an
image file (or if it contains just one volume), or whether the phase encoding
information *varies* between volumes within the image series.
Fixed phase encoding
....................
In the case where both the phase encoding *direction* and the EPI *total readout
time* are equivalent for all volumes within an image, this information is encoded
within two fields: "``PhaseEncodingDirection``" and "``TotalReadoutTime``". These
fields are consistent with `BIDS
`_ (the Brain Imaging Data Structure).
``PhaseEncodingDirection`` can take one of six values: "``i``", "``i-``", "``j``",
"``j-``", "``k``", "``k-``". These correspond to the first, second and third axes of
the corresponding image. Exactly how these axes map to "real" / "scanner" space
axes (i.e. left-right, posterior-enterior, inferior-superior) depends on the
:ref:`transform` that is stored in the respective image header, which may
potentially have been modified internally by *MRtrix3* on image load (discussed
further in the :ref:`non_axial_acquisitions` section). Here, we begin with the most simple
use case:
Take an image that conforms to the RAS (Right-Anterior-Superior) convention common
to both NIfTI and *MRtrix3*. If the phase encoding is applied ``A>>P``
(anterior-posterior), this is the *second* spatial axis, or axis "``j``". However,
the phase encoding is also *reversed* along that axis ("RAS" indicates that voxel
positions along the second axis *increase* when moving toward the anterior of the
brain, whereas "``A>>P``" indicates the opposite). Hence, in this example,
``PhaseEncodingDirection`` would have the value "``j-``".
.. NOTE::
The phase encoding direction is defined specifically with respect to *image
axes*. It is therefore *not* affected by the image *strides*, which only affect
how the data for these axes are arranged when they are stored as a
one-dimensional list of values within a file.
.. NOTE::
The phase encoding direction does *not* relate to "``x``", "``y``" and
"``z``" axis directions in "real" / "scanner" space, as do other
representations of orientation information in *MRtrix3*. This is because phase
encoding specifically affects the appearance of the image *along the image axis
in which phase encoding was applied*. The mapping from image axes to real-space
axes is determined by the image header transform.
To demonstrate this: Imagine if an EPI image were to undergo a rigid-body
rotation. The EPI field inhomogeneity distortions would *still align* with the
relevant image axis after this rotation; but the effective ``x-y-z`` direction
of phase encoding in real-space would change.
``TotalReadoutTime`` provides the total time required for the EPI readout train.
Specifically, this is the time between the centre of the first echo, and the centre
of the last echo, in the train; this is consistent with BIDS, and is sometimes
referred to as the "FSL definition", since it is consistent with relevant
calculations performed within FSL tools. It should be defined in seconds.
Variable phase encoding
.......................
If the phase encoding direction and/or the total readout time varies between
different volumes within a single image series, then the two key-value fields
described above are not sufficient to fully encode this information. In this
situation, *MRtrix3* will instead use a key-value entry "``pe_scheme``" (similar to
the "``dw_scheme``" entry used for the diffusion gradient scheme).
This information is stored as a *table*, where each row contains the phase encoding
direction and the readout time for the corresponding volume; the number of rows in
this table must therefore be equal to the number of volumes in the image. In each
row, the first three numbers encode the phase encoding direction, and the fourth
number is the total readout time. The direction is specified as a unit direction in
the image coordinate system; for instance, a phase encoding direction of A>>P would
be encoded as ``[ 0 -1 0 ]``.
.. _non_axial_acquisitions:
Non-axial acquisitions
----------------------
When images are acquired in such a manner that the axes do not approximately
correspond to RAS convention (i.e. first axis increases from left to right, second
axis increases from posterior to anterior, third axis increases from inferior to
superior), *MRtrix3* will automatically alter the axis :ref:`strides` & transform
in order to make the image *appear* as close to an axial acquisition as possible.
This is briefly mentioned in :ref:`transform` section. The behaviour may
also be observed by running ``mrinfo`` with and without the
``-config RealignTransform false`` option, which temporarily disables this behaviour.
Because phase encoding is defined with respect to the image axes, any
transformation of image axes must correspondingly be applied to the phase encoding
data. When the phase encoding information is stored within the image data,
*MRtrix3* should automatically manipulate such phase encoding information in order
to maintain correspondence with the image data.
Where management of such information becomes more complex and prone to errors
is when it is included in the sidecar information of a JSON file, e.g. as is
commonly now utilised alongside NIfTI images such as in the BIDS format. This
becomes more complex at both read and write stages, each in their own complex way.
1. When *reading* a JSON file, *MRtrix3* will take the transformation that was
applied to the corresponding input image, and apply that to the phase encoding
information.
This has two curious consequences:
1. Running::
mrconvert image.nii -json_import image.json - | mrinfo - | grep PhaseEncodingDirection
and::
cat image.json | grep PhaseEncodingDirection
*may produce different results*. This is because once imported via the
``-json_import`` option, the phase encoding direction is altered to
reflect *how MRtrix3 interprets the image data*, rather than how they
are actually stored on file.
2. Running::
mrconvert image.nii - | mrconvert - -json_import image.json image.mif
*may produce the incorrect result*. This is because information regarding
the transformation that is applied to the NIfTI image in the *first*
``mrconvert`` call in order to approximate an axial acquisition is
*no longer available* in the *second* ``mrconvert`` call. When the
``-json_import`` command-line option is used, it is interpreted with
respect to the input image *for that command* - which, in the above case,
is an MRtrix piped image to which the axial transformation has
*already been applied* - and so must *always* be used immediately in
conjunction with loading the image with which that JSON file is associated.
2. When *writing* a JSON file, *MRtrix3* will attempt to modify the phase
encoding information in order to conform to the limitations of the output
image format alongside which the JSON file is intended to reside.
Unlike the :ref:`mrtrix_image_formats`, :ref:`nifti_format` do not support
arbitrary image strides. When writing image data with non-trivial strides to
a NIfTI image, *MRtrix3* will reorder the three spatial axes in order to
approximate an axial alignment, performing the corresponding modifications
to the image transform. Any exported JSON file must therefore also have the
same transformation applied.
This has the curious consequence that::
mrconvert input.mif output.mif -json_export output.json
cat output.json | grep PhaseEncodingDirection
and::
mrconvert input.mif output.nii -json_export output.json
cat output.json | grep PhaseEncodingDirection
*may produce different results*. A JSON file *must* only be interpreted in
conjunction with the singular image file alongside which it was generated.
In the case of the MRtrix image format, we suggest relying instead on the
storage of sidecar information within the :ref:`header_keyvalue_pairs`
rather than in such an external file, as it allows *MRtrix3* to apply any
requisite modifications to such data to echo modifications to the image
without user intervention.
.. NOTE::
This concept also has consequences for the ``dwifslpreproc`` script when manually
providing the phase encoding direction. The axis and sign of phase encoding
provided to the script must reflect the direction of phase encoding *after*
*MRtrix3* has performed this transformation, i.e. as it is read by any
*MRtrix3* command or as it appears in ``mrview``, *not* the actual encoding
of axes within the file.
Manipulation of phase encoding data
-----------------------------------
The primary purpose of storing this phase encoding information is to automate the
correction of EPI susceptibility distortions. However this can only occur if the
information stored is not invalidated through the manipulation of the corresponding
image data. Therefore, any *MRtrix3* command that is capable of manipulating the
image data in such a way as to invalidate the phase encoding information will
*automatically* modify this phase encoding information appropriately. This includes
modifying the representation of this information between the fixed and variable
phase encoding cases.
Consider, for instance, a pair of b=0 images, where the first was acquired with
phase encoding direction ``A>>P``, and the second was acquired using phase encoding
direction ``P>>A``::
$ mrinfo AP.mif
******************************
Image: AP.mif
******************************
...
PhaseEncodingDirection: j-
TotalReadoutTime: 0.0575
...
$ mrinfo PA.mif
******************************
Image: PA.mif
******************************
...
PhaseEncodingDirection: j
TotalReadoutTime: 0.0575
...
Now watch what happens when we concatenate these two images together::
$ mrcat AP.mif PA.mif AP_PA_pair.mif -axis 3
mrcat: [100%] concatenating "AP.mif"
mrcat: [100%] concatenating "PA.mif"
# mrinfo AP_PA_pair.mif
******************************
Image: AP_PA_pair.mif
******************************
...
pe_scheme: 0,-1,0,0.0575
0,1,0,0.0575
...
When the two input images are concatenated, *MRtrix3* additionally concatenates the
phase encoding information of the input volumes; since it detects that these are not
consistent between volumes, it stores this information using the ``pe_scheme`` header
entry, rather than ``PhaseEncodingDirection`` and ``TotalReadoutTime``.
The ``mrconvert`` command has a number of additional functionalities that can be used
to manipulate this information:
- The ``-import_pe_table`` and ``-export_pe_table`` options can be used to
import/export the phase encoding information from / to file as a table, i.e. in
the format used for the ``pe_scheme`` header entry described above. Note that even
if all volumes in the image have the same phase encoding direction and total
readout time, these options will still import / export these data in table format.
- The ``-import_pe_eddy`` and ``-export_pe_eddy`` options can be used to
import/export the phase encoding information in the format required by FSL's
``eddy`` tool. The `FSL documentation page `_
describes this format in more detail.
- The ``-json_import`` and ``-json_export`` options can be used to import/export
*all* header key-value entries from/to an external JSON file. This may be useful
in particular for operating within the BIDS specification. There is a caveat here:
If you use the ``-json_export`` option on an image with *fixed* phase encoding,
the ``PhaseEncodingDirection`` and ``TotalReadoutTime`` fields will be written as
expected by BIDS; however if the image contains *variable* phase encoding, then
the ``pe_scheme`` header entry will be written to the JSON file, and this will not
be appropriately interpreted by other BIDS tools.
- The ``-set_property`` option may be useful to *override* these header entries if
they are deemed incorrect by some other source of information.